CARDIOVASCULAR JOURNAL OF AFRICA • Volume 30, No 2, March/April 2019
122
AFRICA
of auscultatory types of devices. The non-auscultatory types
include automated, ambulatory and Doppler ultrasound devices.
Invasive blood pressure monitoring
The invasive (intra-arterial line) blood pressure monitoring
involves placement of a cannula into an artery (usually radial)
with the distal end of the cannula attached to tubing, which is
then connected to a pressure transducer. Infusion of heparinised
saline through the tubing and cannula prevents a blood clot
in the set-up.
60
Both numerical measurements and graphical
recording of the blood pressure are displayed on a monitor
61
in
real time, and this helps with dynamic
62
management of patients.
With each heartbeat, the blood pressure waveform rises during
systole and drops during diastole, and the average mean arterial
blood pressure over one cardiac cycle is indicated numerically.
60
In clinical practice, the intra-arterial device is used in the
management of critically ill patients, many of whom have labile
arterial blood pressures. Such clinically ill pregnant women
include those with septic shock, severe obstetric haemorrhage,
eclampsia and other forms of acute organ failure requiring
resuscitation. Arterial blood gas samples may also be obtained
from the arterial line.
Additionally, the validation of a blood pressure device may be
undertaken using an intra-arterial device as a reference standard.
In fact, the validation of neonatal blood pressure devices may
only be performed using an intra-arterial device as a reference
standard.
63
Non-invasive blood pressure monitoring, on the other hand,
is used in all categories of patients, particularly in non-critically
ill patients.
Non-invasive auscultatory blood pressure-measuring
device
Measurement of blood pressure with an auscultatory device
involves using a functional stethoscope to listen to the sound
produced when blood flows through a partially occluded
brachial artery, as well as the subsequent changes in the sound
prior to non-occlusion of the vessel. The origin of the sound is
not clearly understood
64
but is thought to emanate from one or
both of the following: turbulence to blood flow and stretching
of the arterial wall.
61
In this method of blood pressure measurement (mercury/
aneroid sphygmomanometry), the cuff is inflated on the arm
to occlude the brachial artery and is subsequently deflated. The
radial artery in the wrist is palpated during the cuff inflation,
and the pressure at which the arterial pulsation ceases is noted.
Subsequently, the cuff should be inflated further to increase the
pressure by an additional 20–30 mmHg above the point where
the radial pulse is no longer palpable. A stethoscope is placed
on the antecubital fossa, distal to the cuff, to auscultate the
Korotkoff sound during deflation. The first sound (Korotkoff
phase 1) is heard when the pressure in the cuff begins to allow
blood flow through the artery.
65
The cuff pressure at which Korotkoff phase 1 sound is heard
represents the systolic blood pressure. Direct palpation of the
arterial pulse while the cuff pressure is inflating imprecisely
indicates the systolic blood pressure. The diastolic blood pressure
is denoted by the disappearance of the sound (Korotkoff phase
V). Should the sound not disappear, the pressure at muffling
of the sound (Korotkoff phase IV) denotes the diastolic blood
pressure.
The traditional auscultatory device has a mercury column
and its use is no longer popular in many clinical settings due
to concerns of mercury toxicity. Irrespective, the European
Commission Scientific Committee on Emerging and Newly
Identified Health Risks (SCENIHR) recommends that mercury
sphygmomanometers may be used for the validation of other
blood pressure-measuring devices.
66
An aneroid device is a good replacement for the mercury
sphygmomanometers. The aneroid device, however, tends to
lose calibration easily, especially due to mechanical trauma.
Therefore, a mobile aneroid device mounted on a wall or tripod
stand is preferable. Re-calibration of an aneroid device is
required at least once every two years.
6
For research purposes, a
new aneroid sphygmomanometer with a valid calibration status
is recommended for use.
Of note, all blood pressure-measuring devices including
mercury sphygmomanometers require periodic assessment
and recalibration, although the robustness of the equipment
will influence the frequency of this quality check.
67
Evidence-
based details of auscultatory and automated blood pressure
measurement techniques are described in Table 1.
Automated blood pressure-measuring device
The automated or oscillatory blood pressure monitor functions
on the principle that flow of blood through an artery partially
constricted with a cuff causes the arterial wall to vibrate. Such
vibration does not occur if there is no arterial wall constriction.
To explain, the inflation of the cuff of the oscillatory blood
pressure monitor causes partial or complete constriction of the
artery. A cuff pressure above the systolic blood pressure causes
complete constriction of the arterial wall. A reduction in the
cuff pressure just below the systolic blood pressure causes partial
constriction of the arterial wall. With each cardiac contraction, a
pulsatile blood flow occurs and causes vibrations in the partially
constricted arterial wall.
78
The air in the cuff conducts and
transfers the vibration to the transducer in the monitor. The
arterial wall vibrations generate pressure pulses of approximately
3 mmHg in the cuff.
79
It is the transducer that then generates an
electric signal using the transmitted vibrations.
The oscillatory monitor may be classified as inflationary or
deflationary. An inflationary device detects the oscillation in the
arterial wall when the cuff is inflating.
80
It prevents the need for
the pneumatic cuff to be over-inflated substantially above the
systolic blood pressure. In the deflationary type, the cuff inflates
to a pre-determined pressure above the systolic blood pressure
before deflating at a rate of approximately 4 mmHg per second
to detect the arterial wall oscillations.
78
As soon as the pressure
in the cuff is below the diastolic blood pressure, the transmission
of oscillometric pulses ceases because the arterial wall will no
longer be constricted. The pressure pulses are subjected to a
proprietary algorithm of the monitor to generate the blood
pressure reading.
79
Importantly, the pressure at maximum oscillation corresponds
to the mean arterial pressure, while the changes in the oscillatory
amplitude are used to establish the systolic and diastolic blood
pressures.
81
Automated devices therefore measure the systolic and